Gold Standard Phantom's recommendation

Gold Standard Phantoms recommends the following scanner parameters when using the Veriflux phantom for performing fMRI stability analysis.

Acquisition type

EPI

Magnetic Field Strength

3T

Head coil

Coil used for fMRI. Recommendation: 32 or 64-elements coil.

Scan plane

As per clinical protocol, either parallel to AC-PC line or plain axial / transverse.

Field of view

200mm

Slices

As many as possible within a TR = 2000ms, typically 32.

Slice thickness

4mm + 1mm gap

TR

2000ms

TE

30ms

Flip angle

90 degrees

Parallel imaging factor

0 / Off

Matrix

64x64

Number of volumes

200

Scan time

6min 40sec

Dummy scans*

Min 3

Reconstruction

Sum-of-square

*As part of the protocol, you can set up this parameter, which means that the first three scans are going to be acquired but not saved. This brings the magnetization into a steady-state.

Possible alternative protocols

The following set of parameters have been successfully used with Veriflux

Acquisition type

EPI

Scanner Magnetic Field Strength

3T

Head coil

32 coil

Scan plane

Axial

Field of view

200mm

Slices

16

Slice thickness

4mm + 1mm gap

TR

1000ms

TE

30ms

Flip angle

90 degrees

Parallel imaging factor

0 / Off

Matrix

64x64x16

Number of volumes

400

Scan time

6 min 40s

Dummy Scans

Min 3

Reconstruction

Sum-of-square

FBIRN protocol parameters

The following set of parameters are compliant with the FBIRN protocol. The use of 1.5T scanners is supported, but not recommended.

Acquisition type

EPI or Spiral GRE

Scanner Magnetic Field Strength

1.5T

3T

7T

Head coil

Coil used for fMRI. Recommendation: 32 or 64-elements coil.

Scan plane

Axial

Field of view

220mm

Number of Slices

28

32

38

Slice thickness

4mm + 1mm gap

TR

2000ms

TE

40ms

30ms

25ms

Flip angle

90 degrees

Bandwidth

+-100kHz

Parallel imaging factor

0 / Off

Matrix

64x64

Number of volumes

200

Scan time

6min 40sec

Dummy Scans

Min 3

Reconstruction

Sum-of-square