Scanning Protocol

Gold Standard Phantom's recommendation

Gold Standard Phantoms recommends the following scanner parameters when using the Veriflux phantom for performing fMRI stability analysis.

Acquisition type

GRE-EPI, FFE-EPI

Magnetic Field Strength

3T

Head coil

Coil used for fMRI. Recommendation: 32 or 64-elements coil.

Scan plane

Plain axial / transverse

Field of view

200mm

Slices

As many as possible within a TR = 2000ms, typically 32.

Slice thickness

4mm + 1mm gap

TR

2000ms

TE

30ms

Flip angle

90 degrees

Parallel imaging factor

0 / Off

Matrix

64x64

Number of volumes

200

Scan time

6min 40sec

Dummy scans*

Min 3

Reconstruction

Sum-of-square

*As part of the protocol, you can set up this parameter, which means that the first three scans are going to be acquired but not saved. This brings the magnetization into a steady-state.


Possible alternative protocols

The following set of parameters have been successfully used with Veriflux

Acquisition type

GRE-EPI, FFE-EPI

Scanner Magnetic Field Strength

3T

Head coil

32 coil

Scan plane

Plain Axial / Transverse

Field of view

200mm

Slices

16

Slice thickness

4mm + 1mm gap

TR

1000ms

TE

30ms

Flip angle

90 degrees

Parallel imaging factor

0 / Off

Matrix

64x64x16

Number of volumes

400

Scan time

6 min 40s

Dummy Scans

Min 3

Reconstruction

Sum-of-square

FBIRN protocol parameters

The following set of parameters are compliant with the FBIRN protocol. The use of 1.5T scanners is supported, but not recommended.

Acquisition type

GRE-EPI or Spiral GRE

Scanner Magnetic Field Strength

1.5T

3T

7T

Head coil

Coil used for fMRI. Recommendation: 32 or 64-elements coil.

Scan plane

Axial

Field of view

220mm

Number of Slices

28

32

38

Slice thickness

4mm + 1mm gap

TR

2000ms

TE

40ms

30ms

25ms

Flip angle

90 degrees

Bandwidth

+-100kHz

Parallel imaging factor

0 / Off

Matrix

64x64

Number of volumes

200

Scan time

6min 40sec

Dummy Scans

Min 3

Reconstruction

Sum-of-square